Biosafety Cabinet (70%-30%)
- Home
- Biosafety Cabinet (70%-30%)
Our Products Categories
Products Details
A specialized cabinet providing a 70-30 airflow split, offering controlled protection for both operators and the surrounding environment while handling biohazardous materials.
| Direction of Flow | Vertical down flow with 70-30% Exhaust |
| Noise Level | Less than 67 Db on scale ‘A’ When Ambient is not greater than 55 Db on scale ‘A’ |
| M.O.C | Complete body made out of Stainless steel (Make:- Jindal Stainless Limited) |
| Working Perforation Sheet | SS 304 Work table with perforations at front & back with can be removed occasionally fo cleaning purpose |
| Front Door | 1-piece fully openable sliding toughned glass with counter weight arrangement |
| Supply Filter | Minipleat HEPA filter with hot melt technology, which conforms to EU-14 Grade at supply position with an efficiency rating better than 99.999% for 0.3 micron (Make:-RACPE) |
| Pre – Filter | Pre – filter are made from Non- Woven synthetic with HDPE mesh; conforms to EU-06 Grade, with an efficiency rating better than 95% down to 5 micron (Make:- RACPE) |
| Motor Blower Assembly | Motor blower designed of Statically” and dynamically balanced for supply of efficient capacity and static pressure to take care of air flow requirement for entire life of HEPA. (Make:-National Product) |
| Exhaust HEPA Filter | Miniplate HEPA filter with hot melt technology, which conforms to EU-14 Grade at exhaust position with an efficiency rating better than 99.999% for 0.3 micron. (Make:- RACPE) |
| Different Pressure Gauge | Magnehelic Gauge ( 0-25mm:2Nos) to monitor pressured top across HEPA filter & work area with respect to ambient (Make:-Nextech) |
| Lighting | Fluorescent Light (250 Lux) at work surface. (Make:-Philips) |
| ON/OFF Switches | SS ON/OFF Switches for motor & lights |
| Door Opening Alarm | Alarm triggers in case front in case front door is raised more than 8” ( i.e. safe clear door opening during operation ) |
| Exhaust Duct | SS 304 1 ft long elbow duct for exhaust |
| Electrical supply | 230V AC1- Ø 50 HZ/ 41 5V AC 3-Ø 50 HZ |
| Side panels | Double walled flushed glass side panels |
| Vibration Level | Minimum |